Table 5-5-21: RS-232 Calibration Messages
Message Description
START ZERO CALIBRATION Beginning IZS zero calibration
NOX
1
=xxxxx PPB
2
NO
1
=xxxxx PPB
2
NO2
1
=xxxxx PPB
2
FINISH ZERO CALIBRATION
Finished IZS zero calibration
START SPAN CALIBRATION Beginning IZS span calibration
NOX
1
=xxxxx PPB
2
NO
1
=xxxxx PPB
2
NO2
1
=xxxxx PPB
2
FINISH SPAN CALIBRATION
Finished IZS span calibration
START MULTI-POINT CALIBRATION Beginning multi-point calibration
NOX
1
=xxxxx PPB
2
NO
1
=xxxxx PPB
2
NO2
1
=xxxxx PPB
2
FINISH MULTI-POINT CALIBRATION
Finished multi-point calibration
1
Depends on software options installed.
2
Depends on which units are currently selected.
5.5.6 DIAGNOSTIC Commands and Messages
When Diagnostic mode is entered from the RS-232 port, the diagnostic mode issues additional
status messages to indicate which diagnostic test is currently selected. Examples of Diagnostic
mode messages are:
D DDD:HH:MM IIII ENTER DIAGNOSTIC MODE
D DDD:HH:MM IIII EXIT DIAGNOSTIC MODE
Example of turning on the Ozone Generator via the RS-232 port:
D ENTER SIG
D OZONE_GEN=ON
D EXIT