Preparing the Bair Paws Model 875 Warming Unit for Use
Before you begin using the warming unit, make sure that it is safely placed on a at, hard
surface such as a table, or securely mounted on a wall, IV pole, suction wall mount adapter,
or bedrail.
CAUTION: Do not place the warming unit on a soft or uneven surface, such as a bed; this
may cause the air intake to become blocked, compromising the unit’s performance.
A wall-mount bracket is shipped with every warming unit; the brackets for mounting the
warming unit to an IV pole, bedrail, or a suction wall mount adapter are available separately.
Please contact your local sales representative or call 3M patient warming customer service at
1-800-228-3957 for more information about ordering mounting brackets and adapter.
e following instructions for mounting the warming unit assume that the brackets have
already been attached to the wall or to the warming unit.
Placing the Warming Unit on the Wall-mount Bracket
1. Snap the warming unit into position by sliding the clip on the back of the warming unit
into the groove of the wall-mount bracket. e assembly will click when the clip is fully
seated in the wall-mount bracket.
2. Lift the warming unit straight up to remove it from the wall-mount bracket.
N: It may be necessary to gently strike the case bottom in an upward fashion with
the palm of your hand to release the clip from the wall-mount bracket.