Toshiba LF414 Automobile Parts User Manual


 
- 31 -
6F8A0870
CAUTION
Install a switch and fuse to isolate the
LF414/LF600F, LF610F and LF414 from
mains power.
Turn off mains power before conducting wiring
work.
DO
Power supply from mains power can
cause electric shock or circuit
break-down.
DO
Wiring while power is applied can cause
electric shock.
Do not work on piping and wiring with wet
hands.
Ground the LF600F, LF610 independently from
power equipment. (100 ohm or less ground
resistance)
DON’T
Wet hands may result in electric
shock
DO
Operating this product without grounding
can cause electric shock or malfunction.
Do not conduct wiring work with bare hands. For the power supply wiring and grounding wiring,
use crimping terminals with insulated sleeve.
DON’T
Remaining electric charge even if
power is turned off can still cause
electric shock.
DO
There is a risk of electric shock due to
drop-off or loosing, and a risk of fire and
equipment trouble due to heat generation.
Do not modify or disassemble the LF600F,
LF610F, LF602F and LF612F
unnecessarily.
DON’T
Modifying or disassembling this
product can cause electric shock,
malfunction of or damage to this
product.
The label shown left is placed near the
power supply terminal on the
converter.
Be alert to electric shock.
Flowmeter accuracy may be affected by the way wiring is executed. Proceed with correct wiring taking
the precautions in following pages.