Teledyne API Model T360/T360M Operation Manual Troubleshooting and Service
231
Toggles parameter
ON/OFF
Exit returns to
DIAG display & all values
return to software control
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG I/O 0 ) EXT_ZERO_CAL=ON
PREV NE X T JUMP PRNT EXIT
DIAG I/O 22) WHEEL_HTR=ON
PREV NEXT JUMP ON PRNT EXIT
DIAG I/O 28) SAMPLE_PRESSURE=2540 MV
PREV N EXT JUMP PRNT EXIT
DIAG I/O 22) WHEEL_HTR=OFF
PREV NEXT JUMP OFF PRNT EXIT
If parameter is an
input si
nal
If parameter is an output
si
nal or control
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
Figure 8-2: Example of Signal I/O Function
8.1.4. Internal Electronic Status LEDs
Several LED’s are located inside the instrument to assist in determining if the
analyzer’s CPU, I
2
C bus and relay board, GFC wheel and the sync/demodulator
board are functioning properly.