4.6.
Program Monitor Window
When the run has started, you can observe the progress of the program by selecting the
Monitoring option from the Main Menu.
In addition to the information displayed for a Steady State ran, FORMAT displays the
time into program run, number of program cycles left and the Time Hold mode.
The Time Hold mode can be Soak (if theGuaranteed Soak option was enabled at the
Program Run window and the temperature and humidity are not within the preset limits),
Eexternal (through a signal external to the controller, if fitted) or Manual. Manual Time
Hold is effected by pressing the F2 key. The effect is to instantly freeze the program
run time until it is pressed again.
You can stop a program run before it is completed by pressing the F10 key.
Condition : °C
%RH
Status:
Logging:
Program:
Running nroqram
Control:
Not Named
r
Monitor Chamber
Set Point Temperature
Set Point Humidity
Soak
Direct Cooling
Indirect Cooling
Atomiser
| disabled |
|
Vapour Generator
Time Hold
I High RH
none
Not Used
Day Hour Min Sec Cycles Left
[^[^LL][OL] i
5
Not Used
VFT
F1 I Help I F2 I Hold I F3 I Soak I Escl Close |F10 Ishut Down
Fig 8. The Monitor Chamber window
4.7.
Holding a program
To hold a program press F2 from the Monitor Chamber window.
The effect of holding the program will be immediately apparent as the time elapsed count
(timebase) will stop. When a program run is held the temperature and humidity levels
are controlled at the values of the calculated setpoints at the time the hold was initiated,
digital channel output status is also held as determined at that point in the program.
To remove the hold press F2 again, and the program timebase will count again from
where it was held.
40