A SERVICE OF

logo

44 MS6000-EU-EN V1.4 8/12
Overtime Trigger: In Pulse Width trigger mode, it may take some time for a trigger to occur. Since a
complete pulse width is not needed to trigger the oscilloscope, it may be desired to trigger just upon the
overtime point. This is called Overtime Trigger. Press on TRIG to enter Trigger mode.
Options Settings Comments
Type O.T.
Source (F2)
CH1
CH2
Select channel to source the trigger
Polarity (F3)
Positive
Negative
Select to trigger on positive or negative pulses.
Mode (F4)
Normal
Auto
Single
Select the type of trigger. The Normal mode is best for most
applications.
Page (F5) Change page from 1 to 2
Overtime (F2)
(menu page 2)
Adjust timing using the Multifunction control.
Coupling (F3)
(menu page 2)
AC
DC
HF Reject
LF Reject
Noise Reject
Selects the components of the trigger signal applied to the
trigger circuitry.
50% (F4)
Page (F5) Change page from 2 to 1