A SERVICE OF

logo

Test Sample Chromatograms
Test sample chromatograms
229
Detector Type FIDw/MUG
Temp 250 DEGREES C
Inlet Type Ded On-Col Cap
Oven Track On
Temp N/A C
Operating Mode N/A
Purge Time On N/A min
Purge Time Off N/A
min
Oven Temp Temp Programmed (1
ramp)
Init Temp 60
DEGREES C
Init Time 0.5
min
Ramp
Rate 20
Fin Temp 185
Fin Time 1
Flow Param (EPP)
Constant Flow Off
Range 4
COLUMN:
Part No. 19095(#100)
Dimensions 530 MID; 5 M
Sta Phase Methyl
Silicone
START
.331
.112
4.743
5.398
6.
22
.267
FLOW RATES
Carrier (He) 15 +/• 1 ml/min
Hydrogen 30 +/• 1 ml/min
Air 400 +/• 20 ml/min
Makeup (N2) 20 +/• 1 ml/min
Split Vent N/A ml/min
Septum Purge 5 +/• 1 ml/min
SAMPLE:
Type FID Sample
Inj Volume 1
l
Part No. 18710•60170
Composition 0.03%(V/V)each:
C14,C15, and C16 normal paraffin
hydrocarbons in hexane.
C14
C15
C16
HP 5890 Test Sample Operating
Conditions
FID-On-Column Capillary Inlet
Figure 10-5.