A SERVICE OF

logo

Test Sample Chromatograms
Test sample chromatograms
234
Detector Type NPD w/MUG)
Temp 220
DEGREES C
Inlet Type Ded On-Col
Cap
Oven Track On
Temp N/A C
Operating Mode N/A
Purge Time On min
Purge Time Off
min
Oven Isothermal
Init Temp 170
DEGREES C
Init Time 5.0
min
Ramp
Rate 0
Fin Temp
Fin Time
Flow Param (EPP)
Constant Flow Off
Range 0
COLUMN:
Part No. 19095S(#100)
Dimensions 530 MID; 5 M
Sta Phase Methyl Silicone
START
IF
IF
IF
.896
1.966
3.
56
ST
<P
FLOW RATES
Carrier (He) 15 +/• 1 ml/min
Hydrogen 4 +/• 1 ml/min
Air 90 +/• 10 ml/min
Makeup (He) 10+/•1
ml/min
Split Vent N/A
ml/min
Septum Purge 5+/•1
ml/min
SAMPLE:
Type NP-FID Sample
Inj Volume 3
l
Part No. 18789•60060
Composition 0.65 ppm azobenzene,
100 ppm octadecane, 1.0 ppm
malathion in isoioctane
AZOBENZENE
OCTADECANE
MALATHION
HP 5890 Test Sample Operating
Conditions
NPD-On-Column Capillary Inlet
Figure 10-10.