Section 2: Installation Considerations
September 2013
Installation considerations
Figure 2-12. Rosemount 5300 Series with remote housing
When the remote housing is used together with the Rosemount 5300 Series, there may be a
limit on measuring range and a decrease in accuracy. This is because the remote housing
introduces a double bounce in the tank signal at 1.5 times the remote housing cable length.
Because the double bounce occurs at approximately 1.5 times the remote housing cable length,
a longer remote housing cable can increase the maximum measuring range for distances up to
10 ft (3 m). Ta bl e 2 -7 shows the maximum recommended measuring range with remote
housing for different remote housing lengths, installation types, dielectric constants, and probe
types.
Table 2-7. Rosemount 5300 Series remote housing measuring range
Dielectric
Constant
Rigid Single
8 mm
Rigid Single
13 mm
Flexible Single
1 m
Remote
Housing
Chamber / pipe
installations 4 in.
(100 mm)
2
80
2
80
10 ft (3 m)
(1)
10 ft (3 m)
4 ft (1.25 m)
10 ft (3 m)
(1)
(1) Accuracy may be affected up to +
1.2 in. (30 mm).
15 ft (4.5 m)
(1)
15 ft (4.5 m)
(1)
4 ft (1.25 m)
10 ft (3 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
4 ft (1.25 m)
159 ft (48.5 m)
(1)
(2) Required chamber/pipe size is 3 or 4 in. (75 -100 mm).
Tank installations
2 m
Remote
Housing
Chamber / pipe
installations 4 in.
(100 mm)
2
80
2
80
10 ft (3 m)
(1)
10 ft (3 m)
9 ft (2.75 m)
10 ft (3 m)
(1)
15 ft (4.5 m)
(1)
15 ft (4.5 m)
9 ft (2.75 m)
10 ft (3 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
9 ft (2.75 m)
154 ft (47 m)
(1)
Tank installations
3 m
Remote
Housing
Chamber / pipe
installations 4 in.
(100 mm)
2
80
2
80
10 ft (3 m)
15 ft (4.5 m)
15 ft (4.5 m)
14 ft (4.25 m)
15 ft (4.5 m)
(1)
33 ft (10 m)
(1)
(2)
33 ft (10 m)
(1)
(2)
14 ft (4.25 m)
149 ft (45.5 m)
(1)
Tank installations